Ir-192 HDR sources
Monte Carlo results for high dose rate 192Ir brachytherapy sources.
Introduction
This website presents the dosimetric characterization of four high dose rate brachytherapy source models. Data are available for the GammaMed Plus and VariSource 2000 sources, distributed by Varian Medical Systems, and for the MicroSelectron mHDR-v2 source and its new version, mHDR-v2r, distributed by Nucletron BV.
The results have been obtained by Monte Carlo simulations with the PENELOPE 2008.1 code, following the recommendations of both the AAPM Task Group TG-43 report and its update, TG-43 U1.
Results are available for the radial dose function, the 2D anisotropy function, the air kerma strength and the absorbed dose rate constant, as well as the absorbed dose rate tables in water, both for calculations in a semi-infinite phantom and in a spherical phantom of 15 cm radius.
Available sources
Simulation details
In this work, version 2008.1 of the PENELOPE code has been used for the Monte Carlo simulation of radiation transport. This code complies with the recommendations of the AAPM Task Group 43 report and its update, TG-43 U1. Among other features, it includes a set of numerical data and analytical fits from the XCOM and EPDL97 libraries to evaluate the photoelectric effect cross sections.
On the one hand, to calculate the air kerma strength (Sk), the fluence of photons crossing a sphere of 1 m radius centered on the active core of the source has been obtained by Monte Carlo simulations, considering a geometry with the source in vacuum and scoring the photons according to their energy and to the polar angle at which they cross the surface. The discretization used for the surface was ΔΘ = 1º. The energy of the photons detected at the surface has been scored in a histogram with 1 keV bins, to avoid possible artifacts due to the use of wider cells.
On the other hand, to calculate the dose rate in water, the source has been considered immersed in a semi-infinite water phantom (ρ = 0.998 g/cm3), with the origin of the coordinate system at the center of the iridium core and the Z axis along the source, so that cylindrical symmetry is preserved. Separate simulations have also been performed in a finite phantom of 15 cm radius, in order to compare under the same conditions with other published data. The positive direction of the Z axis has been taken towards the tip of the capsule. Two different spatial discretizations have been used: a cylindrical one, with annular cells of Δz = 0.05 cm and Δr = 0.05 cm, and one in spherical coordinates, with cells of Δr = 0.05 cm and ΔΘ = 1º. The simulation parameters and cutoff energies have been set to ensure that electrons with energies below the absorption energy have a range smaller than half of any dimension of the energy scoring cells (in the continuous slowing down approximation). With the data obtained, a 2D table of the dose rate in water has been built, and the dose rate constant, the radial dose function and the 2D anisotropy function have been calculated following the AAPM TG-43 protocol for the line source approximation. The active length considered for each source model was 3.5 mm for the GammaMed Plus and MicroSelectron mHDR-v2r sources, 3.6 mm for the MicroSelectron mHDR-v2 source and 5 mm for the VariSource 2000 source.
Ir-192 emission spectrum
Ir-192 is a nuclide produced by neutron bombardment of metallic iridium, with a half-life of 73.831(8) days1, which decays by beta emission (with probability 95.24(4)%) and electron capture (with probability 4.76(4)%). The emission of characteristic X-rays and the de-excitation of the decay products give rise to the emission spectrum shown in the following figure. An emission of 2.3575 photons per decay has been considered. These values have been obtained from the LUND/LBNL nuclear data search engine. The electrons emitted directly in the nuclear decay have not been simulated, since their contribution to the calculations is much smaller than the uncertainties in the definition of the geometry, in the Ir-192 distribution within the source or in the cross sections used by the simulation codes.
Other data for these sources
References
- [1]Nath R, Anderson L, Luxton G, Weaver K, Williamson J and Meigooni A Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43, Med Phys 1995;22:209-34.
- [2]Rivard M, Coursey B, DeWerd L, Hanson W, Saiful Huq M, Ibbott G, Mitch M, Nath R and Williamson J, Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations, Med Phys 2004;31:633-74.
- [3]Li Z, Das R, DeWerd L, Ibbott G, Meigooni A, Pérez-Calatayud J,Rivard M, Sloboda R and Williamson J, Dosimetric prerequisites for routine clinical use of photon emitting brachytherapy sources with average energy higher than 50 keV, Med Phys 2007;34:37-40.
- [4]Daskalov G, Löffler E and Williamson J, Monte-Carlo aided dosimetry of a new high dose-rate brachytherapy source, Med Phys 1998;25:2200-2208.
- [5]Granero D, Vijande J, Ballester F and Rivard M, Dosimetry revisited for the HDR 192Ir brachytherapy source model mHDR-v2, Med Phys 2011;38:487-94.
- [6]Ballester F, Puchades V, Lluch J, Serrano-Andrés M, Limami Y, Pérez Calatayud J and Casal E, Technical note: Monte-Carlo dosimetry of the HDR 12i and Plus 192Ir sources, Med Phys 2001;28:2586-91.
- [7]Angelopoulos A, Baras P, Sakelliou L, Karaiskos P and Sandilos P, Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source, Med Phys 2000;27:2521-2527.
- [8]http://www.stal.com.cn/e_technicaldata.asp.
- [9]Chu S, Ekström L and Firestone R, The Lund/LBNL Nuclear Data Search, Version2.0, Feb 1999, available at http://nucleardata.nuclear.lu.se/NuclearData/toi/
- [10]Borg J and Rogers D, Monte Carlo calculations of photon spectra in air from Ir-192 sources, NRC Report PIRS-629R, 1998.
- [11]Salvat F, Fernández-Varea J and Sempau J, PENELOPE-2008: A code system for Monte Carlo simulation of electron and photon transport, Workshop Proceedings, Barcelona, Spain. ISBN 978-92-64-99066-1.
- [12]Born M, Atomic Physics. Blackie and Son, London.
- [13]Hubbell J, Veigele W, Briggs E, Brown R, Cromet D, Howerton R, Atomic form factors, incoherent scattering functions, and photon scattering cross sections. J Phys Chem Ref Data 1975;4:471-538.
- [14]Cullen D, Hubbell J and Kissel L, EPDL97 The evaluated photon data library,’97 version, Report UCRL-50400 vol.6, rev.4, parts A and B (Livermore,CA: Lawrence Livermore National Laboratory).
- [15]Ribberfors R, X-ray incoherent scattering total cross sections and energy-absorption cross sections by means of simple calculation routines. Phys Rev A 1983;27:3061-70.
- [16]Berger M and Hubbell J, XCOM:Photon Cross Sections on a Personal Computer, Report NBSIR 87-3597 (Gaithersburg,MD: NIST) ,1987.
- [17]ICRU (1998). Fundamental Quantities and Units for Ionising Radiation. Report No. 60, International Commission on Radiation Units and Measurements, 7910 Woodmont Ave., Bethesda, Maryland 20814.
- [18]Higgins P, Attix F, Hubbell J, Seltzer S, Berger M and Sibata C, Mass Energy-Transfer and Mass Energy-Absorption coefficients, including in-flight positron annihilation for photon energies 1 keV to 100 MeV. NISTIR 4680, National Institute of Standards and Technology, Gaithersburg, MD (1991).
- [19]Hubbell J and Seltzer S, Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients (version 1.03). http://physics.nist.gov/xaamdi [2004]. NIST, Gaithersburg, MD.
- [20]Taylor R and Rogers D, An EGSnrc Monte Carlo-calculated database of TG-43 parameters, Med Phys, 2008;35,4228-4241.
- [21]Taylor R and Rogers D, EGSnrc Monte Carlo calculated dosimetry parameters for 192Ir and 169Yb brachytherapy sources, Med Phys 2008;35:4933-4944.
- [22]Lis Boada M, Ballester Pallarés F and Pérez Calatayud J, Parámetros Dosimétricos de fuentes de Cs-137 e Ir-192 en Braquiterapia, http://www.uv.es/braphyqs.
- [23]Berenguer Serrano R, Rivera Giménez M, Núñez Quintanilla A and Gutiérrez Pérez M, Evaluación de la dosimetría de una fuente de Ir-192 de alta tasa en un medio con dispersión incompleta mediante cálculo de Monte Carlo, Revista de Física Médica 2006;7(3):107-112.
- [24]Casado F, García-Pareja S, Cenizo E, Mateo B, Bodineau C and Galán P, Dosimetric characterization of an 192Ir brachytherapy source with the Monte Carlo code PENELOPE, Phys Med, 2010;26(3):132-139.
- [25]Evaluation of measurement data - Guide to the expression of uncertainty in measurement, International Organization for Standardization (ISO). Joint Committee for Guides in Metrology (JCGM 100, 2008). Corrected version 2010. http://www.bipm.org/utils/common/documents/jcgm/JCGM_100_2008_E.pdf
- [26]DeWerd L, Ibbott G, Meigooni A, Mitch M, Rivard M, Stump K, Thomadsen B and Venselaar J, A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO. Med Phys, 2011;38:782-801.
- [27]Pérez-Calatayud J, Granero D and Ballester F, Phantom size in brachytherapy source dosimetric studies, Med Phys 2004;31:2075-81
- [28]Melhus C and Rivard M, Approaches to calculating AAPM TG-43 brachytherapy dosimetry parameters for 137Cs, 125I, 192Ir, 103Pd, and 169Yb sources, Med Phys 2006;33:1729-37
- [29]Chandola RM, Tiwari S, Kowar MK, Choudhary V. Monte Carlo and experimental dosimetric study of the mHDR-v2 brachytherapy source. J Can Res Ther 2010;6:421-6.
- [30]Chandola RM, Tiwari S, Painuly NK, Choudhary V, Azad SK, Beck M. Monte Carlo study of dosimetric parameters and dose distribution effect of inhomogeneities and source position of GammaMed Plus source. J Can Res Ther 2011;7:29-34
1 Throughout this website, values in parentheses give the uncertainty in the last significant digit; read 73.831(8) as 73.831 ± 0.008.